8-19
IM AQ6370C-17EN
Program Function
1
2
3
4
5
6
7
8
App
Index
CENTER
Program Command
Description
Instrument models on which the
command is valid, parameter ranges,
and supported variables.
CENTER WL ****.***nm Sets measurement center
wavelength.
AQ6370C/AQ6370D
600.000 to 1700.000nm (0.001 step)
AQ6373/AQ6373B
350.000 to 1200.000nm (0.001 step)
AQ6375/AQ6375B
1200.000 to 2400.000nm (0.001 step)
CENTER WL
@@@@@
Sets the value of variable
@@@@@ to measurement
center wavelength
@@@@@: E
1
, G, H, I, J, K, O
1
, P, Q,
R, S, T
1
, U
1
, V
1
, X, Y, Z, WM, W1, W2,
W(CH), PKWL, MEANWL, WDMWL(CH),
NFWL(CH)
On the AQ6370C or AQ6370D
valid variables
WAM1, WAM2, WAM3, WAM4
CENTER FREQ ***.****
THz
Sets measurement center
frequency.
AQ6370C/AQ6370D
176.5000 to 500.0000THz (0.0001 step)
AQ6373/AQ6373B
250.0000 to 856.5000THz (0.0001 step)
AQ6375/AQ6375B
125.0000 to 250.0000THz (0.0001 step)
CENTER FREQ
@@@@@
Sets the value of variable
@@@@@ to measurement
center frequency
@@@@@: E
1
, G, H, I, J, K, O
1
, P, Q, R,
S, T
1
, U
1
, V
1
, X, Y, Z, WM, W1, W2,
WAM1, WAM2, WAM3, WAM4, W(CH),
PKWL, MEANWL, WDMWL(CH),
NFWL(CH)
CENTER WNUM
****.***cm
-1
Sets measurement center
wavenumber.
AQ6375/AQ6375B
4167.000 to 8333.000cm
-1
(0.001 step)
CENTER WNUM
@@@@@
Sets the value of variable
@@@@@ to measurement
center wavenumber
@@@@@: G, H, I, J, K, P, Q, R, S,
X, Y, Z, WM, W1, W2,W(CH), PKWL,
MEANWL
START WL ***.**** nm
Sets measurement-starting
wavelength.
AQ6370C/AQ6370D
50.000 to 1700.000 nm (0.001 step)
AQ6373/AQ6373B
1.000 to 1200.000 nm (0.001 step)
AQ6375/AQ6375B
600.000 to 2400.000 nm (0.001 step)
START FREQ ***.****
THz
Sets measurement-starting
frequency.
AQ6370C/AQ6370D
11.5000 to 500.0000 THz (0.0001 step)
AQ6373/AQ6373B
10.0000 to 856.5000 THz (0.0001 step)
AQ6375/AQ6375B
60.0000 to 250.0000 THz (0.0001 step)
START WNUM
****.***cm
-1
Sets measurement-starting
wavenumber.
AQ6375/AQ6375B
2067.000 to 8333.000cm
-1
(0.001 step)
STOP WL ****.***nm
Sets measurement-ending
wavelength.
AQ6370C/AQ6370D
600.000 to 2250.000 nm (0.001 step)
AQ6373/AQ6373B
350.000 to 1625.000 nm (0.001 step)
AQ6375/AQ6375B
1200.000 to 3000.000 nm (0.001 step)
STOP FREQ ***.****THz Sets measurement-ending
frequency.
AQ6370C/AQ6370D
176.5000 to 6650.0000 THz (0.0001 step)
AQ6373/AQ6373B
250.0000 to 999.9000 THz (0.0001 step)
AQ6375/AQ6375B
125.0000 to 315.0000 THz (0.0001 step)
STOP WNUM
****.***cm
-1
Sets measurement-ending
wavenumber.
AQ6375/AQ6375B
4167.000 to 10433.000cm
-1
(0.001 step)
8.3 Program Function Commands