7-85
IM AQ6370C-17EN
Remote Commands
1
2
3
4
5
6
7
8
App
Index
:SENSe:SWEep:SPEed
Function
Sets/queries the sweep speed.
Syntax
:SENSe:SWEep:SPEed<wsp>1x|2x|0|1
:SENSe:SETTing:FCONnector?
1x|0: Standard
2x|1: Twice as fast as standard
Response 0 = 1x, 1 = 2x
Example
:SENSE:SWEEP:SPEED 2x
:SENSE:SWEEP:SPEED? -> 1
Explanation • This is a sequential command.
• This command can be used on the AQ6370C/
AQ6370D/AQ6373B/AQ6375B.
:SENSe:SWEep:STEP
Function
Sets/queries the sampling interval for
measurements.
Syntax
:SENSe:SWEep:STEP<wsp><NRf>[M]
:SENSe:SWEep:STEP?
<NRf> = The sampling interval for measurement
[m]
Example
:SENSE:SWEEP:STEP 1PM
:SENSE:SWEEP:STEP?-> +1.00000000E-012
Explanation • When the function of automatically setting
the sampling interval for measurement
(
SENSe:SWEep:POINts:AUTO
command)
is ON, the sampling number to be measured
that has been set can be queried.
• When the function of automatically setting
the sampling number to be measured
(
SENSe:SWEep:POINts:AUTO
command) is
ON, this command will be automatically set to
OFF.
• When the sampling interval for measurement is
set using this command, the sampling intervals
for measurements (
SENSe:SWEep:POINts
)
will be automatically set.
• This is a sequential command.
:SENSe:SWEep:TIME:0NM
Function
Sets/queries the time taken from the start to the
end of measurements when measurement is
made in the 0-nm sweep mode.
Syntax
:SENSe:SWEep:TIME:0NM<wsp><integer>
[SEC]
:SENSe:SWEep:TIME:0NM?
<integer> = Measurement time [sec] (0 =
MINIMUM)
Example
:SENSE:SWEEP:TIME:0NM 10SEC
:SENSE:SWEEP:TIME:0NM? -> 10
Explanation This is a sequential command.
:SENSe:SWEep:TIME:INTerval
Function
Sets/queries the time taken from the start of a
sweep to that of the next sweep when repeat
sweeps are made.
Syntax
:SENSe:SWEep:TIME:INTerval<wsp><int
eger>[SEC]
:SENSe:SWEep:TIME:INTerval?
<integer> = Measurement time [sec] (0 =
MINIMUM)
Example
:SENSE:SWEEP:TIME:INTERVAL 100sec
:SENSE:SWEEP:TIME:INTERVAL? -> 100
Explanation This is a sequential command.
:SENSe:SWEep:TLSSync
Function
Sets/queries the synchronous sweep function.
Syntax
:SENSe:SWEep:TLSSync<wsp>OFF|ON|
0|1|
:SENSe:SWEep:TLSSync?
Response 0 = OFF, 1 = ON
Example
:SENSE:SWEEP:TLSSYNC ON
:SENSE:SWEEP:TLSSYNC? -> 1
Explanation This is a sequential command.
• With the AQ6370D/AQ6373B/AQ6375B, the
command is invalid.
:SENSe:WAVelength:CENTer
Function
Sets/queries the measurement condition center
wavelength.
Syntax
:SENSe:WAVelength:CENTer<wsp><NRf>[
M|HZ]
:SENSe:WAVelength:CENTer?
<NRf> = Measurement center wavelength [m]
Response
<NRf> [m|Hz] (AQ6370C/AQ6370D/AQ6373/
AQ6373B)
<NRf> [m|Hz|m
-1
] (AQ6375/AQ6375B)
Example
:SENSE:WAVELENGTH:CENTER 1550.000NM
:SENSE:WAVELENGTH:CENTER?->
+1.55000000E-006
Explanation • For the AQ6375/AQ6375B, to set using the
wavenumber, do not add any units when
in Wavenumber mode, and just enter the
numerical value.
• This is a sequential command.
:SENSe:WAVelength:SPAN
Function
Sets/queries the measurement condition
measurement span.
Syntax
:SENSe:WAVelengthSPAN<wsp><NRf>
[M|HZ]
:SENSe:WAVelength:SPAN?
<NRf> = Measurement span [m]
Response
<NRf> [m|Hz] (AQ6370C/AQ6370D/AQ6373/
AQ6373B)
<NRf> [m|Hz|m
-1
] (AQ6375/AQ6375B)
Example
:SENSE:WAVELENGTH:SPAN 20.0NM
:SENSE:WAVELENGTH:SPAN?->
+2.00000000E-008
Explanation • For the AQ6375/AQ6375B, to set using the
wavenumber, do not add any units when
in Wavenumber mode, and just enter the
numerical value.
• This is a sequential command.
7.6 Instrument-Specific Commands