PROTOC 200
What to do if...
7 - 53
Note
Manually carry out zeroing and calibration after troubleshooting. Press
the ALARM key to shut off the sensor alarm.
7.3
Unstable readings
– Leakage in gas supply
– See section 7.1 T
HE
DISPLAY
RESPONDS
SLOWLY
OR
NOT
AT
ALL
TO
THE
CALIBRATION
SOLUTION
OR
SAMPLE
Cause
Actions
Calibration/zeroing deviates from
the nominal values by more than
the specified hysteresis.
– Carrier gas flow fluctuates
– Use the appropriate
pressure regulator and
adjust it correctly
– The displacement in the U-tube
is not constant
– Ensure that the flow of liquid
through the system is steady
– Check the gas outlet section
and the sparge vessel for a
steady flow of sample
through the system
– Correct the positions of the
PTFE tubes if necessary
– Leakage in gas supply
– See section 7.1 T
HE
DISPLAY
RESPONDS
SLOWLY
OR
NOT
AT
ALL
TO
THE
CALIBRATION
SOLUTION
OR
SAMPLE
Cause
Actions
AL sensor is
displayed