PROTOC 200
Installation
3 - 27
3.6
Checklist for commissioning the equipment
Check the following points before beginning measurement:
l
Is reagent solution available (not older than 1 month)?
l
Is fresh standard solution available for calibration?
l
Is cleaning solution available?
l
Is distilled water present in the U-tube?
l
Is carrier gas available at a pressure of 2.2 bar?
3.7
Commissioning
1
Carry out the preparatory steps as described in the checklist.
2
Switch on the main power supply to the PROTOC 200 (if it was
turned off).
3
Plug in the controller power cable.
4
Switch on the IR detector (if it was turned off).
5
Open the gas inlet.
6
Switch on the pump.
7
Check that liquids and gas are flowing.
8
Switch on the UV lamp.
9
Allow one hour for the analyzer to stabilize.
Switching on the
equipment