Setup
PROTOC 200
5 - 46
Proceed as follows to change the controller settings:
5.5.2
Setting the date/time
5.5.3
Configuring the analog output
Set the analog output to current (4–20 mA) or voltage (0–10 V) as fol-
lows:
1
Press the ENTER/MENU key. The MAIN MENU appears on
the display.
2
Press F4. The CONTROLLER SETUP menu appears. The
hour setting under TIME/DATE is highlighted.
3
Set the hour with the UP/DOWN keys.
4
Press F1 again. The minutes setting is highlighted.
5
Set the minutes with the UP/DOWN keys.
6
Press F1 again. Follow the same procedure to set the date
(month, day, year).
7
Return to measurement: Press the ENTER/MENU key repeat-
edly until the measurement display appears.
1
Press the ENTER/MENU key. The MAIN MENU appears on
the display.
2
Press F4. The CONTROLLER SETUP menu appears.
3
Press F2. The analog output setting is highlighted.
4
Set to 4–20 mA output current or 0–10 V output voltage with
the UP/DOWN keys.
5
Return to measurement: Press the ENTER/MENU key repeat-
edly until the measurement display appears.