PROTOC 200
Setup
5 - 39
5.2.7
Calibration
The equipment automatically calibrates 0 to 2 times a day. The calibra-
tion duration can be selected within the range of 5 to 20 minutes, de-
pending on the analyzer response time and application demands.
Calibration can also be started manually. The actual calibration is car-
ried out in the last 3 seconds of the calibration time frame. During cali-
bration, the last signal is frozen on the analog output.
Proceed as follows to change the duration and number of calibrations
per day:
You can now manually start calibration:
1
Press the ENTER/MENU key. The MAIN MENU appears on
the display.
2
Press F1. The SENSOR SETUP menu appears.
3
Press F4. The EXTENDED SETUP menu appears.
4
Press F1. The CONFIGURE AUTO MODE menu appears.
5
Press F3. The number of calibrations over a 24 hour period is
highlighted.
6
Set the required number (0 ... 2) with the UP/DOWN keys.
7
To lengthen or shorten the calibration duration, press F3 fol-
lowed by the UP or DOWN key.
8
Press F3 again. MANUAL is highlighted.
9
Start the calibration with the UP or DOWN key. The time dis-
play counts down.
Calibration /
CALIBRATE
Starting calibration
manually
Summary of Contents for PROTOC 200
Page 1: ...PROTOC 200 TOC Analyzer ba43200e01 11 00 Operating Manual...
Page 10: ...Safety PROTOC 200 2 10...
Page 26: ...Installation PROTOC 200 3 26 Figure 3 9 Carrier gas valve 3...
Page 28: ...Installation PROTOC 200 3 28...
Page 54: ...What to do if PROTOC 200 7 54...
Page 59: ...PROTOC 200 Appendix 9 59 9 Appendix 9 1 Flow diagram...
Page 60: ...Appendix PROTOC 200 9 60 9 2 Socket connections for the analog output...