Setup
PROTOC 200
5 - 40
5.2.8
Purging
After each calibration, cleaning and zeroing operation, the equipment
automatically purges the system before updating the display and out-
puts. The purging duration can be selected within the range of 5 to 15
minutes. The last signal is frozen on the analog output during all purg-
ing cycles.
Proceed as follows to change the duration and number of purging cy-
cles per day:
You can now manually start a purging cycle:
1
Press the ENTER/MENU key. The MAIN MENU appears on
the display.
2
Press F1. The SENSOR SETUP menu appears.
3
Press F4. The EXTENDED SETUP menu appears.
4
Press F1. The CONFIGURE AUTO MODE menu appears.
5
Press F4 twice. The purging duration is highlighted.
6
Set the required number (5 ... 15) with the UP/DOWN keys.
7
Press F4 once. MANUAL is highlighted.
8
Start the purging cycle with the UP or DOWN key. The time dis-
play counts down.
PURGE
Starting a purge cycle
manually