Setup
PROTOC 200
5 - 48
5.5.5
Serial interface (option)
For controllers equipped with a serial interface, the following settings
can be made in the COMMUNICATIONS menu:
l
Baudrate
l
Printer on or off
l
Address
l
Data report every 4 or 24 hours
Configuring the
interface
1
Press the ENTER/MENU key. The MAIN MENU appears on
the display.
2
Press F4. The CONTROLLER SETUP menu appears.
3
Press F4. The baudrate is highlighted.
4
Set the value (300, 600, 1200, 2400, 4800 or 9600) with the
UP/DOWN keys.
5
Press F2. The printer setting is highlighted (PRINTER ON or
OFF).
6
Set the printer to ON or OFF with the UP/DOWN keys.
7
Press F3. The address is highlighted.
8
Set the address (0 ... 9; A ... F) with the UP/DOWN keys.
9
Press F4. The report setting is highlighted.
10
Set the report output to 24 hours (24 HR), 4 hours (4 HR) or to
OFF.
11
Return to measurement: Press the ENTER/MENU key repeat-
edly until the measurement display appears.