PROTOC 200
Setup
5 - 47
5.5.4
Setting the output voltage/current range
Proceed as follows to finely adjust the range limits of the current or volt-
age output to adapt the output to different loads.
1
Press the ENTER/MENU key. The MAIN MENU appears on
the display.
2
Press F4. The CONTROLLER SETUP menu appears.
3
Press F3. The lower limiting value for the current output is high-
lighted.
4
Set the value (3.50 ... 4.50 mA) with the UP/DOWN keys.
5
Press F3 again. The upper limiting value for the current output
is highlighted.
6
Set the value (18.00 ... 22.00 mA) with the UP/DOWN keys.
7
Press F3 again. Set the lower limiting value (-0.50 ... + 0.50 V)
and the upper limiting value (9.00 ... + 11.00 V) for the voltage
output as described above.
8
Return to measurement: Press the ENTER/MENU key repeat-
edly until the measurement display appears.