PROTOC 200
Setup
5 - 33
5
Setup
5.1
Controller function
Overview of the
operation
l
The controller receives a signal from the analyzer and displays it as
a measurement.
l
Automatic functions such as zeroing, calibration and cleaning are
time-controlled.
l
Two alarm points are available that can be adjusted by the user and
one sensor alarm is available that indicates analyzer malfunctions.
l
Pressing the alarm button during an automatic function reduces the
function duration to 35 s (Attention: Inaccurate data may result if the
analyzer is calibrating or zeroing at that moment.).
l
The controller only accepts data at the end of zeroing or calibration
if the signal lies /- 10 % of the nominal value. Otherwise, an
error message is generated. Measuring continues despite the error
message.
Controller times and functions are set in the main menu as follows:
1
Starting point: Measurement display.
2
Press the MENU key. The MAIN MENU appears. The various
setup menus can be accessed from this menu.
Accessing the main
menu