PROTOC 200
Setup
5 - 37
5.2.5
Cleaning
The equipment automatically carries out a cleaning cycle 0 to 4 times
a day. The duration is set to 3 minutes. Cleaning can also be started
manually. The last signal is frozen on the analog output during all clean-
ing cycles.
Proceed as follows to change the duration and number of cleaning cy-
cles per day:
You can now manually start a cleaning cycle:
1
Press the ENTER/MENU key. The MAIN MENU appears on
the display.
2
Press F1. The SENSOR SETUP menu appears.
3
Press F4. The EXTENDED SETUP menu appears.
4
Press F1. The CONFIGURE AUTO MODE menu appears. The
number of cleaning cycles over a 24 hour period is highlighted.
5
Set the required number (0 ... 4) with the UP/DOWN key.
6
To lengthen or shorten the cleaning cycle duration, press F1
followed by the UP or DOWN key.
7
Press F1 again. MANUAL is highlighted.
8
Start the cleaning cycle with the UP or DOWN key. The time
display counts down.
CLEANING
Starting a cleaning
cycle manually