PROTOC 200
Setup
5 - 43
5.3.3
Setting up the access code
PROTOC 200 settings can be protected by a code function that uses a
3-digit access code. The equipment is delivered with the code function
deactivated.
Attention
Please write down the code if you activate the code function or change
the code. The MAIN MENU in which the settings are made is not ac-
cessible without the code. If you should forget the code, please contact
the WTW service department.
Proceed as follows to activate the code function or to change the code:
1
Press the ENTER/MENU key. The MAIN MENU appears on
the display.
2
Press F2. The ALARMS SETUP menu appears.
3
Press F4. The access code setting is highlighted.
4
Set as required with the UP/DOWN keys (ON or OFF).
5
Press F4 again. The first digit of the 3-digit access code is high-
lighted.
6
Set as required with the UP/DOWN keys (F1, F2, F3 or F4).
7
Press F2. Set the values for the remaining two digits as de-
scribed above.
8
Return to measurement: Press the ENTER/MENU key until the
measurement display appears.