DOC: 03-933001-01
DESC:
Assy, Basic Electrical, (120v Shown)
PAGE:
REV
17 of 23
1
PREPARED BY: Sax Makthepharack REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
07 - 18 - 00
Dave Maderos
1.8
1177 INJECTOR INSTALLATION
1.8.1
Obtain the appropriate 1177 injector, 03-925996-01 (for 120VAC, Type 21), 03-925996-02 (for 230VAC, Type 21) and 03-925996-03 (for
101VAC, Type 21) or 03-925997-01 (for 120VAC, Type 25), 03-925997-02 (for 230VAC, Type 25) and 03-925997-03 (for 101VAC, Type
25).. Attach injector switch cable assembly, 03-933510-01 to the micro switch of the 1177 injector. Connect the BROWN wire to the center
tab of the switch and BLACK to the remaining tab.
1.8.2
With the proper clearance in the column oven insulation, install the 1177 injector onto the top of the oven chassis using two each 8-32 X 1/
4" screws, 12-901165-00 as shown. Install Test Injector Cover, 03-933013-01 to insure proper alignment of 1177 Injector. If required,
loosen injector mountings screws and adjust accordingly.
1.8.3
Route the 1177 injector heater, RTD and injector switch wires through the keyhole slot in the right side of the chassis. Place a split
grommet, 24-799873-00 around the 1177 and column temperature probe wires and snap into chassis top.
1.8.4
Bundle the 1177 injector and column probe wires together and secure to chassis top using one each Nylon Clamp, 22-119944-00 and one
each 8-32 x 1/4 Screw, 12-901165-00.
Continue Next Page