DOC: 03-933001-01
DESC:
Assy, Basic Electrical, (120v Shown)
PAGE:
REV
22 of 23
1
PREPARED BY: Sax Makthepharack REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
07 - 18 - 00
Dave Maderos
1.11
FINAL WIRING/CONNECTIONS
1.11.1 Obtain electronics module assembly, 03-933006-01 and set into position next to the rear of instrument. Starting from the bottom, plug the
circuit breaker connector into J19 and the transformer primary connector into J18. Plug the column oven heater connector into J17 and the
column fan motor connector into J16. Plug the injector heater connector into J13 and the transformer secondary connector into J15. Plug
the Door/LED /stepper motor 18-pin connector into J6 and the EFC flat cable connector into J3. Plug the power fail solenoid connector into
J21 and the Column/injector probe connector into J1.
Continue Next Page
COLUMN PROBE/INJECTOR PROBE
EFC FLAT CABLE (J3)
POWER FAIL SOLENOID
CONNECTOR (J21)
DOOR/LED AND STEPPER
MOTOR 18-PIN (J6)
TRANSFORMER
PRIMARY CONNECTOR (J18)
TRANSFORMER SECONDARY
CONECTOR (J15)
CIRCUIT BREAKER CONNECTOR (J19)
COLUMN HEATER CONNECTOR (J17)
COLUMN FAN MOTOR (J16)
INJECTOR HEATER
AND SWITCH CONNECTOR (J1)
03-933006-01
ASSY, ELECTRONICS MODULE
CONNECTOR (J13)