DOC: 3900GC TEST STATION
DESC: 3900GC GALILEO TEST
PAGE:
REV
11 of 12
1
PREPARED BY: Sax Makthepharack REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
10 - 10 - 00
Bob Jolley
1.5.3
Verify initial values:
A) Pressure Set point and Actual Pressure equilibrate to 10psi.
B) Oven temp. equilibrates to 50 deg.
C) Injector temp. equilibrates to 100 deg.
D) Split state and ratio are off.
E) No faults exist.
1.5.4
At the end of stabilization, tap the 3900 injector switch to begin the
run.
1.5.5
During the run verify the following:
A) During first minute.
I.
Oven vent closes completely during oven heat-up.
II.
Oven heats up to and stabilizes at 100 deg.
III. Split state on, split ratio 10, Total Flow increases.
IV. Pressure rises towards 22.5.
B) During second minute.
I.
Pressure stabilizes at 22.5.
II.
Oven vent opens completely during oven cool down.
III. Oven cools to and stabilizes at 50 deg.
IV. Split ratio changes to 100, Total Flow increases.
C) After 2 minutes into run split ratio changes to 20, Total Flow
decreases.
I.
At end of run. Split state and ratio are off, Total Flow
decreases.
1.5.6
Open oven door and verify oven turns off.
1.5.7
Verify 3900 faults, by clicking on .
CLICK HERE
STEP 1.4.7