DOC: 3900GC TEST STATION
DESC: 3900GC GALILEO TEST
PAGE:
REV
12 of 12
1
PREPARED BY: Sax Makthepharack REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
10 - 10 - 00
Bob Jolley
1.5.8
Set Injector (C) to 90 and click Close & Update.
1.5.9
Verify 3900 faults, and red LED illuminates on 3900.
1.5.10 Click and return the Injector (C) to 325 deg and then
click Close & Update.
1.5.11 Exit System Control program, by clicking the
File
menu and then
clicking
Exit
.
1.5.12 Turn off 3900 and helium. Disconnect power, computer cable and
helium line.
1.5.13 Remove injector column fitting cap and uncap and reconnect
septum purge line.
CLICK HERE
STEP 1.5.10
CLICK HERE
STEP 1.5.8