DOC: 03-933016-21/25
DESC: EFC Sub-Assembly, Installation
PAGE:
REV
5 of 5
1
PREPARED BY: Sax Makthepharack REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
09 - 25 - 00
Sax
Figure 13
Figure 14
1.1.15 Lay the EFC sub-assembly down as shown in Figure 13 to measure and adjust tubing length. Adjust the SST tube that is not
labeled to 16" +/- .25" and the other three tubes (two tubes for type 25) to 13" +/- .25". Tighten tie wrap to secure tubes.
1.1.16 Set the assembly upright and gather the four tubes and install a final tie wrap on the bracket as shown in Figure 14.
1.1.17 Route the clear vent tube through the obround hole in the bracket, leave approximately 6" of tubing protruding through bracket not
shown.