DOC: 03-933001-01
DESC:
Assy, Basic Electrical, (120v Shown)
PAGE:
REV
13 of 23
1
PREPARED BY: Sax Makthepharack REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
07 - 18 - 00
Dave Maderos
1.6
DOOR VENT/VENT DUCT INSTALLATION
1.6.1
Make all the proper bends in intake duct, 03-933196-01 per template.
1.6.2
Place intake-duct around vent door shaft and secure first with 2 each 8-32 X 1/4" screws, 12-901165-00 into exhaust vent duct,
03-933189-01 as shown.
1.6.3
Snap nylon bearing sleeve, 21-128961-00 into vent shaft bracket 03-933162-01 as shown.
1.6.4
Secure vent shaft bracket and intake duct to rear oven chassis using three each 8-32 X 1/4" screws, 12-901165-00 as shown.
1.6.5
Weave intake duct arm support into notched opening as shown. Loosely secure support to intake duct using one each 8-32 X 1/4" screws,
12-901165-00 and one each cable clamp, 22-119785-00 as shown.
1.6.6
Install appropriate column fan motor assembly, 03-933012-01 (for 120VAC), 03-933012-02 (for 230VAC) and 03-933012-03 (for 101VAC)
to channel, 03-933170-01 using four each 8-32 X 1/4" screws, 12-901165-00. Motor wires should be located closest to the line filter.
NOTE:
Verify there are 2 each vibration isolators supporting the bottom of the fan motor assembly.