DOC: 03-933001-01
DESC:
Assy, Basic Electrical, (120v Shown)
PAGE:
REV
19 of 23
1
PREPARED BY: Sax Makthepharack REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
07 - 18 - 00
Dave Maderos
1.9
FINAL WIRING/CONNECTIONS
1.9.1
Obtain 4-pin connector housing, 51-406927-00 and install the 1177 heater contacts into pins 2 and 4 as shown. NOTE: There is no
polarity requirement.
1.9.2
Install the contacts from the 1177 injector probe into pin 1 and 11 and the injector switch cable wires into pins 10 (BLACK) and pin 20
(BROWN) on the 20-pin connector attached to the column oven temperature probe assembly, 03-925375-03.
1.9.3
Place the six wires from the Door/LED harness assembly, 03-933509-01 into the 18-pin connector on the stepper motor assembly,
03-925306-01. Attach wires as follows: Pin 4 - WHT/BRN, Pin 5 - BRN, Pin 6 - BLK, Pin 13 - GRN, Pin 14 - Red, Pin 16 - BLK. NOTE:
There is no polarity requirement for black wires. Reference wiring diagram below.
1.9.4
Install two each tie wraps, 22-119650-00 around the Door/LED and stepper motor harnesses. Place one approximately 2 inches and one
approximately 6 inches from the 18-pin connector.
1.9.5
Obtain two each cable clamps, 22-120322-00. Snap one clamp into the hole in the exhaust duct, 03-933189-01 and the other into the hole
on the bottom side of the chassis top, as shown. Obtain three each standoff cable clamp, 22-300001-00. Snap one each into the center
hole of channel, 03-933170-01. Snap the other two into the bottom of the chassis top as shown.
Continue Next Page
* WHT/BLK = 1
10 = WHT/BLU *
* BLU = 2
11 = RED *
* WHT = 3
12 = BLK *
WHT/BRN = 4
13 = GRN
BRN = 5
14 = RED
BLK = 6
15
7
16 = BLK
8
17
9
18
WIRING DIAGRAM FOR STEP 1.9.3
NOTE: * WIRES PREVIOUSLY INSTALLED