6.8
Communication Specifications
59
6
6.8
Communication Specifications
6.8.1 RS-232C
6.8.1.1 Transfer Specifications
6.8.1.2 Connector Specifications
6.8.1.3 CABLE
(1) Cable connection
Use the connection cable as shown in Fig. 6-5 to connect the turbo molecular pump and
computer. (Number on figure is pin number of connector.)
(2) Cables used
For connection with this connector, the communication cable with shield is necessary.
The shield of the cable should be connected to case.
Interface
RS-232C
Synchronous system
Asynchronous
Transmission rate
9600 bps (fixed)
Character configuration
Start bit : 1
Data bits : 8
Parity : None
Stop bit : 1
Flow control
None
Connector
SERIAL connector (shared with RS-485)
Connector type
Circular type connectors
Pin assignment
1 : RD (Received data)
2 : SD (Sent data)
3 : SG (Signal ground)
* Do not connect any other pins except for the above-mentioned.
Summary of Contents for UTM1600 Series
Page 2: ...This page is intentionally left blank...
Page 10: ...viii Location of Label...
Page 26: ...12 2 2 4 External I F Panel This page is intentionally left blank...
Page 32: ...18 3 3 3 Controller This page is intentionally left blank...
Page 38: ...24 4 4 4 Standards Fulfilled This page is intentionally left blank...
Page 54: ...40 5 5 5 Notes on Transportation This page is intentionally left blank...
Page 78: ...64 6 6 8 Communication Specifications This page is intentionally left blank...
Page 79: ...7 7GAS PURGE...
Page 88: ...74 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Page 101: ...10 10WARRANTY CLAUSES...
Page 104: ...90 10 This page is intentionally left blank...
Page 106: ...Index Index 2 This page is intentionally left blank...