- 33 -
Mode No. Preset output:Setting of OUT1
6
A B C D E
6.0 0 0 0
Hysteresis operation
0:Invalidity
1:Effective
Output mode
0:Comparison
1:Holding
Up limit/Lower limit selection
0:Up limit
1:Lower limit“Immediately”
2:Lower limit“Delay”
Output indicated selection
0:Functional stop
1:Rate meter
2:Ratio
It functions by the comparative result of the indicated value and preset value.
Please refer to “ How to setthe preset value”(pages 47-48) for the setting of the
preset value.
[ Hysteresis operation ]
0:Invalidity
Hysteresis operation is not carried out.
1:Effective
Hysteresis operation is carried out eith the preset value of "Mode №5"
[ Output mode ]
0:Comparison
It outputs, when an indicated value becomes more than a preset value or the
following.
It will become the output OFF if an indicated value comes outside a setting
range.
1:Holding
It outputs, when an indicated value becomes more than a preset value or the
following.
If it outputs once,an output will be held until reset input is turned on.
[ Up limit/Lower limit selection ]
0:Up limit
Select that this is output when the upper limit is exceeded.
1:Lower limit“Immediately”
Serect that this is output when the lower limit is exceeded.
2:Lower limit“Delay”
Select that this is output when the lower limit is exceeded from the state
Upper limit exceeded once.
[ Output indicated selection ]
0:Functional stop
The function of a OUT1“preset output”is suspended.
1:Rate meter
Rate meter value is compared with a preset value.
2:Ratio
Ratio value is compared with a preset value.
Summary of Contents for LV1000 Series
Page 16: ...12 The setting menu Each mode transition Preset value setting...
Page 17: ...13 Mode setting Scaling setting...
Page 18: ...14 Linearization setting 8 step preset value setting...
Page 19: ...15 Test mode...
Page 57: ...53 Analog adjustment mode flow chart...
Page 64: ...60 Dimensions Fig 26...