Page 56 of 101 pages
5.1.2 Temp 2:
Temperature of sensor 2.
Refer to Temp1 for instructions.
5.1.3 Temp 3 (Option):
Temperature of sensor 1.
Refer to Temp1 for instructions.
5.1.4 Temp 4 (Option):
Temperature of sensor 2.
Refer to Temp1 for instructions.
5.1.5 Chm Pres: (Chamber Pressure)
This screen requires a code to be entered.
Press the UP key once, to enter the value ØØ1 and press ENTER.
This provides the screen giving 4 pressure values:
A100
R100
A300
R300
Pressing the right button (F3) allows you to pass to the different values, which
may be changed using the UP/DN key only.
After F3 is pressed when it is on the last value, it moves on to the next screen.
Pressing the middle button (F2) reads the value of the sensor.
Pressing STOP will stop the cycle.
5.1.6 Jkt Pres: (Jacket Pressure)
Jacket Pressure.
Refer to Chm Prs for instructions.
5.1.7 Gen Pres:
Generator Pressure.
Refer to Chm Prs for instructions.
5.1.8 D1 Pres:
Pressure in Gasket1.
Refer to Chm Prs for instructions.
5.1.9 D2 Pres: (Option)
Pressure in Gasket 2.
Refer to Chm Prs for instructions.
After Enter is pressed, the Calibration options are concluded and the display
returns to the ST.BY screen.
This calibration method enables the technician to change a certain point
without affecting a second point.
There are two working temperatures: 134, 121 . When checking chamber TC
for 134
º
C it shows 132
º
C. When checking chamber TC for 121
º
C it shows
121
º
C