Page 22 of 101 pages
Program: 7
Front door close Msw
PT100 Drain Control
PT100 Condense (dual)
IMT Door Seal -front door
IMT Jacket
IMT Chamber
Open front door
Close front door
Ejector
Buzzer
13 – Cool drain
15 – Water to ejector
43 – Air inlet
53 – Seal release
73 – Fast exh
74 – Slow exh.
75 – Vacuum
92 – Steam to jacket
93 – Steam to chamber
94 – Steam to seal or
34 – Compressed Air
inlet
273ºF / 310kPa
274ºF / ~ 210kPa
160 kPa
230ºF / ~ 115kPa
100kPa
50kPa
25kPa
0kPa
Start (Close door)
Door secure
Temp. control
Pres. control
On (activated)
115kPa
Ex. Mode 1
274ºF - 7min
15min
Condition to open
T<230ºF ; P<115kPa