Page 62 of 101 pages
Air
♦
Connect the air connector on the autoclave to the building air source by
means of a 3/8” pipe at 6 to 7.5 bar (90 to 110 psi)
♦
If there is no external air source, use a small air compressor.
Water
♦
The autoclave should be supplied with tap water; its hardness should not
0.7-2mmol/l. See also
2.4
Water quality.
♦
The tap water pressure should be 4 to 6 bar (50 to 80 psi).
♦
Connect the feed water to the network by a 1/2” pipe.
Drain
Connect the following outlets directly to a drain funnel. Or connect them
through a drain collector pipe of 2” that should be covered. The drain system
should be vented.
Connect the system drain by 3/4 “pipe.
Electricity
Connect the power cord: Three phases, Neutral and Grounding to the live
terminals R, S, T, N in the electrical box and the ground wire- GND to the
grounding screw connection. The required outlet for the electrical supply is,
115 V/ 10 A
.