Page 16 of 101 pages
Program: 3 & 4
Front door close Msw
PT100 Drain Control
PT100 Condense (dual)
IMT Door Seal- door
IMT Jacket
IMT Chamber
Open door
Close door
Ejector
Buzzer
13 – Cool drain
15 – Water to ejector
43 – Air inlet
53 – Seal release
73 – Fast exh
74 – Slow exh.
75 – Vacuum
92 – Steam to jacket
93 – Steam to chamber
94 – Steam to seal or
34 – Compressed Air inlet
273ºF / 310kPa
270/274ºF / ~ 210kPa
160 kPa
230ºF / ~ 115kPa
100kPa
50kPa
25kPa
0kPa
Start (Close door)
Door secure
Temp. control
Pres. control
On (activate
For program 4 - 3 vacuum pulses.
115kPa
Ex. Mode 1
270/274ºF; 5min.
Condition to open
T<230ºF; P<115kPa
15 min