Page 74 of 101 pages
Symptoms
Possible causes check-
ups and tests
Corrections
11. Drying incomplete
packs remain wet.
11.1 Insufficient steam pressure
to jacket.
11.1 Set or fix the
pressure switch.
11.2 Insufficient vacuum.
11.2 See problems
related to vacuum
par. 2 and 4.
11.3 Steam trap does not
eliminate the condensate.
11.3 Clean or repair the
steam trap.
11.4 Incorrect loading of the
autoclave.
11.4 Reload the
autoclave
appropriately.
12. Problems related to
the control system.
12.1 Check if the flat cable or
MOLEX connectors
between the control boards
and keyboard are well
plugged in.
12.1 Plug in firmly the
MOLEX
connectors.
12.2 Power supplies out of order. 12.2 Fix or replace the
power supply.
12.3 Control boards or keyboard. 12.3 Replace
defective
board of
microcomputer
unit.
12.4 Paper to printer missing or
quality unsuitable.
12.4 Insert a paper roll
inside the printer.
Ensure the
appropriate type of
paper is used.
12.5 Flat cable connection
between control board and
printer is not plugged in
correctly.
12.5 Plug in connection
or replace it.
12.6 Printer damaged.
12.6 Replace the
printer.
12.7
Temperature
sensor
disconnected or loose
connection. Sensor
damaged.
12.7 Fix connection or
replace sensor.
12.8 Pressure transducer, bad.
Mechanical or electrical
connections or transducer
damaged.
12.8 Fix connection or
replace transducer.