![background image](http://html1.mh-extra.com/html/tuttnauer/5596-sp-1/5596-sp-1_operation-and-maintenance-manual_1182760072.webp)
Page 67 of 101 pages
9.2 Printer Handling
The printer is driven and controlled automatically by the control
unit, while the autoclave performs a sterilization program.
Figure 1
Figure 2
To set the paper roll in the printer perform the following steps:
1. Gently push the front panel clips. Remove panel.
2. Pull out printer gently.
3. Place roll with printing showing surface. (See Figure 1) Gently
push the paper face down into insertion opening (A) in Figure
2. Keep pressing the feed switch (B) until the paper comes
out from the print head (C).
4. When the paper comes out, insert it in the paper cutter (the
slot in the front panel) and reassemble the front panel on the
unit.
Printer is now ready for use!
NOTE: If the paper is not pulled in by the rollers even when
you press the feed switch (B) push the paper in.
5. To ensure correct operation of the printer perform the
following:
♦
Turn the main switch to the OFF position.
♦
Turn the main switch to the ON position while
pressing the feed switch. Print all built in
characters to verify that the printer performs an
operation.
Check the following to ensure the proper operation of the
printer:
♦
Avoid contact between the paper and the hot parts
of the autoclave, as the paper will be blackened.
♦
Do not pull out the paper roll from the paper
insertion opening.
♦
Use only the 58mm. wide thermal paper rolls,
supplied by your dealer.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com