Page 60 of 101 pages
Unloading
Upon cycle completion, unload the load from the sterilizer immediately;
put it aside to cool down.
Put the carriage or the containers aside to cool down.
Do not transfer hot load to metal shelves for cooling. Perform a visual
inspection to ascertain that sterilizing indicators have made the required
colour change, and that the load is dry.
The load shall be rejected if:
a.
The package has been compressed.
b.
The package is torn.
c.
The load is wet (water droplets signs).
d.
The load fell on the floor.
e.
Condense can be observed on the lid.
f.
The PCD or other Indicator present – Faulty cycle.
During loading and unloading use safety gloves and glasses in
accordance with local safety regulations and good practice.