Heat Step - Waiting time between heating and sterilization
This parameter allows partial entry into the sterilization stage with a reduction of
overshoot and the reduction difference between the different sensors that are
entered into the chamber for control or validation.
In the heating stage the system stops heating before entering sterilization
according to the Heat Step value. When this value is 2 minutes, the system stops
at 35.6ºF before entering sterilization and will proceed at a rate of
33.80 ºF/min during the Heat Step.
Entry code
11
Resolution
1 min.
Minimum value
Ø psia.
Maximum value
10 psia.
This is the final parameter under Code 11. After pressing ENTER the display will
return to the ST.BY screen
Dry Vac. -
Dry Vac. =0 ; Vacuum pump works continuously.
Dry Vac. = 5; Vacuum pump works for 5 min. after which air is introduced for 30
sec. of ON- and 30 sec. OFF intervals (depending on the interval selected by the
operator).
Entry code
11
Resolution
0.1 min.
Minimum value
Ø
Maximum value
13
Code 13 Parameters:
- The addition to the required pressure for sterilization
This parameter allows the pressure point that will be maintained during the
sterilization stage to be set in the following manner:
If the desired temperature is 273ºF, the required pressure (from the tables) is
44.2psi.
When the is Ø, the system will be maintained at 44.2 psi during the
sterilization stage. When the equals 0.7 psi, the system will be
maintained at 44.9 psi in the sterilization stage. It is advised to work with a value
of 0.5 psi.
Entry code
11
Resolution
0.1 psia
Minimum value
0.1 psia
Maximum value
4 psia
Select PT100
There are 2 sensors that regulate cycles – one in the condense water, and one in
drain.
Selections:
1. Sensor on the condensate line - dual (No.1) & (No.2).
2. Sensor on the drain output (No.5).