Page 38 of 101 pages
4.2
Description of Displayed Messages and Safety Measures
03 – Manual Stop
This message is displayed and the FAIL indicator lights up
after the STOP key is pressed for more than 1 second during
the cycle (excluding the drying stage).
100 – Man. Stop
This message is displayed and the FAIL indicator lights up
after the STOP key is pressed for longer than one second in
the drying stage.
05-Door Unlock
This message is displayed if either of the following occurs:
The START button is pressed and the door is not closed;
If during a cycle or the pressure drops in the gasket.
06-Temp. Error
If the temperature sensor, disconnected for any reason during
the heating stage, temperature is either higher than 140ºC or
lower than 5ºC..
07- Low Heat
This message is displayed and sterilization does not start if
the autoclave has not reached sterilization temperature after
heating for the maximum time as defined in the HEAT T.O.
parameter.
08- Low Temp.
This message is displayed, fail indicator lights up and cycle is
aborted, if the temperature drops for more than 5 sec. below
the required sterilization temperature.
09- High Temp.
This message is displayed, fail indicator lights up and
program is aborted: If the temperature rises 3ºC (5ºF) above
the required sterilization temperature during the sterilization
cycle.
10- Low Pres.
This message is displayed, fail indicator lights, and the
program is aborted if the pressure drops for more than 5 sec.
below the pressure correlated to the required sterilization
pressure.
11-High Pres.
This message is displayed, fail indicator lights, and the
program is aborted, if pressure rises above the pressure
correlated to the sterilization temperature plus +3ºC (5ºF)- for
more than 5 sec.
12-Low Vacuum
This message is displayed if after 18 min. the system has not
reached at the Vac Dip1or Vac Dip2 values during the pre-
vacuum stage. It also appears if the Leak Test program fails.
13 – Jacket not Ready
This message is displayed if the START key was activated
before the jacket reached the desired pressure.
21 – Door1 Safe Switch
If while raising door1 of an automatic sliding door, the door1
switch closes the door will open, and this message will
appear.
22 – Door2 Safe Switch
(not applicable)
if during raising door2 of an automatic sliding door the door2
switch closes, the door will open, and this message will
appear.