Page 43 of 101 pages
4.3.3 System Set-Up (changing parameters and values)
Note:
♦
The system allows 15 seconds for values to be entered. If this time elapses,
the display will move on automatically to the next screen.
♦
When codes are not entered within the 15 seconds, the RUN and FAIL lamps
light and the display return to WAIT and then the Stand BY screen.
♦
If the code is incorrect the above will also occur.
♦
When the display shows ØØØØ and a single value (i.e. between 1-9) is to be
entered, the Ø must first be pressed. The single value can not be entered on
its own, for example, to enter the number 4, the value is entered by pressing
the Ø three times and then the number 4.
♦
When a value is to be entered, the right command keypad is Enter:
If the value is entered via the digit keypad, after the final digit is entered, the
system automatically transfers to the next screen, and Enter need not be
pressed.
If the value is entered using the UP/DN arrows, the Enter keypad must always be
pressed to progress to the next screen.
The system setup includes the following functions:
Clock – Date and Time, Language, Doors Number, Door Type & Atmospheric
Pressure
To check the system has been set up correctly, press the OPTION keypad in the
ST.BY screen.
The screen displayed is as follows:
Press the SETUP keypad. The system requests a code.
Enter the code and the display automatically moves on to the next screen after the
last digit is entered.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com