The Gain value must be equal or less than …
The Gain value must be equal or greater than …
The Offset value must be equal or less than …
The Offset value must be equal or greater than …
Description
: These messages are displayed when the user is trying to download
Gain and Offset by pressing the 'Download' button in the 'Calibration' window.
Reason:
Gain or Offset value displayed in Gain or Offset text boxes are not in
permitted limits.
Tips:
The value in the Gain or Offset boxes must be in Min/Max limits.
Invalid Gain Value!
Invalid Offset Value!
Description
: These messages are displayed when the user is trying to download
Gain and Offset by pressing 'Download' button in the 'Calibration' window.
Reason:
Probably the Gain or Offset value displayed in Gain or Offset text boxes
are not numeric data.
Tips:
Type numeric value in Gain or Offset text boxes.
Illegal Port number:
…
Description
: This message is displayed when the user is pressing the 'OK'
button in the 'Options' window.
Reason:
Wrong port number is inserted.
Tips:
Type value: 0>VAL>17 in 'Port Number' text box.
Illegal Baudrate: …
Description
: This message is displayed when the user is pressing 'OK' button in
the 'Options' window.
Reason:
Wrong Baud rate is inserted.
Tips:
Type values 9600 or 19200 in 'Baudrate' text box.
Illegal Wait Loop value: …
Description
: This message is displayed when the user is pressing 'OK' button in
the 'Options' window.
Reason:
Wrong Wait Loop value is inserted.
Tips:
Type numeric value in 'Wait Loop' text box.
Illegal Autoclave Address:
Description
: This message is displayed when the user is pressing 'OK' button in
the 'Options' window.
Reason:
Wrong Address value is inserted.
Tips:
Type numeric value: 0<VAL<17.