Page 41 of 101 pages
4.3.2 Selecting a Cycle
A cycle can only be selected during Stand By from the Stand By / Ready screen
which displays the cycle number and name in the top row.
A cycle can be selected as follows:
♦
Press the UP/DN keypad to change the cycle where pressing UP moves to
the next program and pressing DOWN moves to the previous program.
♦
Press on a number. When pressing on keypad 4 the chosen program will be
number 4, and so on.
Selecting a 2 digit cycle:
When 2 numbers are pressed within 2 seconds of each other, then both the
numbers define the program number (when the program number is 10 or above).
For example; to choose program 12, number 1 must be pressed, and within 2
seconds number 2 must be pressed.
Show Parameters
After selecting the cycle, pressing the SHOW command key presents the following
screen:
where the message displayed in the 3rd row describes the following:
♦
the temperature; the sterilization temperature
♦
the time; the sterilization time
♦
D-Time; drying time.
This screen is displayed for 10 seconds after which it automatically returns to the
ST.BY screen.
This command key is not active for cycle 12 – Leak Test.
Once the following conditions have been fulfilled,
♦
There is enough steam pressure in the jacket.
♦
Enough air pressure for the pneumatic valves (option).
♦
Doors
closed.
♦
Enough water pressure for cooling (option).
♦
Enough water in the vacuum water reservoir.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com