10.4 Absolute Pressure Transducer
The control unit is fitted with three pressure transducers, for the control and
monitoring of chamber pressure, for the control of jacket pressure, and for the
control of gasket pressure.
The transducer type TECSIS 3296 is a membrane pressure sensor and electronic
measuring circuit, having the following specifications:
Pressure gauge: 0-4bar abs.
Output span current: 4-20mA
Supply voltage: 10V to 30V
The terminals at the connector of the device are 1 (+24V) and 2 (OUT) to the
controller output.