Page 47 of 101 pages
Ex Mode – The method for exhausting the steam at the end of the process
Ex Mode
Fast Ex
1
N.A.
2
Slow Ex. (waste)
3
Slow Ex. (liquids)
4
Cooling with compressed air
5
Cooling with water circulation
6
EndTemp – Ending Temperature (Exhaust Stage)
Defines the temperature at which the process ends for opening the door.
Entry code
11
Resolution
34ºF
Minimum value
180ºF
Maximum value
280ºF
PulsNum. – No. of pulses in the pre vacuum stage
Entry code
11
Resolution
1
Minimum value
0
Maximum value
10
VacDip1 – Vacuum value in the first pulse
Defines the vacuum value in pulse no.1 of the pre vacuum stage.
Entry code
11
Resolution
0.1 psia
Minimum value
1 psia
Maximum value
14 psia
VacDip2 – Vacuum value in the remaining pulses
This value defines the vacuum in the remaining pulses in the pre-vacuum stage
(not just pulse no. 2)
Entry code
11
Resolution
0.1 psia
Minimum value
1 psia
Maximum value
14 psia
VacTime1 – Waiting time in the first pulse
In the prevacuum stage, in the first pulse after reaching VacDip1, there is a waiting
time known as VacTime1 (in seconds).
This delay occurs so vacuuming can continue.
Entry code
11
Resolution
1 second
Minimum value
3 seconds
Maximum value
1800 seconds (30 minutes)