Page 39 of 101 pages
23 – Door1 Gasket Low
Vacuum
This message indicates that during opening Door1, you do
not achieve the proper vacuum in Gasket1.
24 – Door2 Gasket Low
Vacuum
(not applicable)
If whilst opening Door2, you do not achieve the proper
vacuum in Gasket2.
25 – Vac Res. Empty
This message is a result of a safety mechanism indicating
that If during the pre-vacuum stage, drying stage, vac. test or
leak test it is revealed that the lower float switch in the
vacuum water reservoir is open for more than 1 second (so
there is no water entering the water reservoir and the pump
has therefore been closed to protect it).i.e. Vfloat –L = “1”
27 – First Close Door
This message appears if there was an attempt to perform
IN/OUT Test while the door was open.
4.2.1 Informative Messages
Calibration
This message appears automatically on panel 2 during
calibration.
Closing Door Wait
When an automatic door is closing, this message is displayed
so that the operator will wait until the closing operation of the
automatic door ends.
Open Door Wait
Whilst an automatic door is opening, this message is
displayed in order that the operator will wait until the opening
operation of the automatic door ends.
Door1 Open
This message appears when Door1 is open. It will also
appear if the door is supposedly closed but there is not
enough pressure in the gasket to seal it properly.
Door2 Open
(not applicable)
This message appears when Door2 is open. It will also
appear if the door is supposedly closed but there is not
enough pressure in the gasket to seal it properly.
Door 1 & 2 Open
(not applicable)
This message appears when both doors are open.
Under Test
This message appears on panel 2 during the IN/OUT Test
and during Set Up.