Learning Center
Measure a Micro-Volume Sample
84 NanoDrop Eight User Guide
Thermo Scientific
2. Lift the instrument arm and clean the upper and lower
pedestals with new laboratory wipe.
3. Measure a blank:
–
Pipette 1–2 µL blanking solution onto the lower
pedestal and quickly lower the arm
–
Click
Blank
and wait for the
measurement to complete
Tip
is On, blank measurement
starts automatically after you lower the arm.
–
Lift the arm and clean all pedestals with a new
laboratory wipe
4. Measure the first sample:
–
Pipette 1-2 µL sample solution onto the pedestal
and quickly lower the arm (see
for more information)
–
Start the sample measurement:
–
if
–
if Auto-Measure is off, lower arm and click
Measure
–
When the sample measurement is completed, the
spectra and reported values are displayed.