Thermo Scientific
Before you begin...
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see
To measure a Protein A280 sample
1. From the Home screen, select the
Proteins
tab and select
Protein A280
.
2. Specify a
if desired.
3. Pipette 1–2 µL blanking solution onto the lower pedestal and lower the arm.
4. Click
Blank
and wait for the measurement to complete.
Tip
: If
is On, the blank measurement starts automatically after you
lower the arm.
5. Lift the arm and clean all pedestals with a new laboratory wipe.
6. Pipette 2 µL sample solution onto the pedestal and lower the arm.
7. Start the sample measurement:
–
–
if Auto-Measure is off, lower arm and click
Measure
.
When the sample measurement is completed, the spectrum and reported values
are displayed (see the next section).
8. When you are finished measuring samples, click
End Experiment
.
9. Lift the arm and clean all pedestals with a new wipe.