8
Maintenance
Decontaminating the Instrument
Thermo Scientific
NanoDrop Eight User Guide 133
Decontaminating the Instrument
Decontaminate the instrument after measurements with samples that contain
and before returning the instrument to us for maintenance or
repair.
Tip
The PR-1 pedestal reconditioning compound is the easiest way to
recondition the pedestals. If you don’t have a PR-1 kit, follow these steps:
1. Lift the instrument arm and pipette 3 µL 0.5M HCl onto the lower pedestals.
2. Lower the arm and wait 2–3 minutes.
3. Lift the arm and clean all pedestals with a new laboratory wipe.
4. Pipette 3 µL DI H
2
O onto the lower pedestals.
5. Lower the arm and wait 2–3 minutes.
6. Lift the arm and clean all pedestals with a new wipe.
NOTICE: Support the instrument arm with one hand while you buff the upper
pedestal to avoid damaging the arm.
7. Fold a clean laboratory wipe into quarters and use it to vigorously buff the
surface of each pedestal at least 50 times.
8. Use canned air to remove any paper residue from the pedestals.
Note
If your instrument requires maintenance or repair,
or your local
distributor.
Note
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will
permanently damage the quartz fiber optic cables.
• To prevent damage from spills, keep containers of liquids away from the
instrument.
• Do not use a squirt or spray bottle on or near the instrument as liquids may
flow into the instrument and may cause permanent damage.
• Do not attempt to remove the diaphragm around the lower pedestals as it is
permanently affixed to the instrument.
• Do not allow HCl, alcohol, bleach, acetone or any other solvent to remain on
the diaphragm for more than one minute or it may loosen the seals. If the
diaphragm becomes loose,
•
Summary of Contents for NanoDrop Eight
Page 6: ...Contents vi NanoDrop Eight User Guide Thermo Scientific...
Page 24: ......
Page 56: ...5 Protein Applications Measure Protein A280 56 NanoDrop Eight User Guide Thermo Scientific...
Page 76: ...6 Custom Applications Measure Custom 76 NanoDrop Eight User Guide Thermo Scientific...
Page 142: ...8 Maintenance 142 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...