Measure a Micro-Volume Sample
Thermo Scientific
NanoDrop Eight User Guide 83
• Use
to ensure proper liquid column formation.
• Use new tip for each blank and sample aliquot.
• Use new aliquot of sample for each measurement.
• When the measurement is complete, open the sampling arm and wipe the
samples from both the upper and lower pedestals using a soft laboratory wipe.
If solvents are used, make sure they are compatible with the pedestals. (see
“Compatible Solvents” in
).
Recommended sample volumes
To measure a micro-volume sample
Application
Sample Volume
Nucleic acid (aqueous solution)
1 µL
a
a
Use 2 µL for samples that contain materials that may reduce surface tension such as a surfactant.
Purified protein
2 µL
Other protein applications such as Bradford or BCA
2 µL
Microbial cell suspensions
2 µL
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will
flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will
permanently damage the quartz fiber optic cables.
1. From the Home screen, select an application from
one of the application categories, such as
UV-Vis
, or
Custom
Methods.
Summary of Contents for NanoDrop Eight
Page 6: ...Contents vi NanoDrop Eight User Guide Thermo Scientific...
Page 24: ......
Page 56: ...5 Protein Applications Measure Protein A280 56 NanoDrop Eight User Guide Thermo Scientific...
Page 76: ...6 Custom Applications Measure Custom 76 NanoDrop Eight User Guide Thermo Scientific...
Page 142: ...8 Maintenance 142 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...