Operation
Settings
3-46
42iQTL Instruction Manual
Thermo Scientific
●
Prereactor Bkg Check Offset:
Displays prereactor concentration based
on the last attempted background calibration. High alarm shows
user defined limit for acceptable background check offset.
●
NO Bkg Check Offset:
Displays NO concentration based on the last
attempted background calibration. High alarm shows user defined
limit for acceptable background check offset.
●
NO
x
Bkg Check Offset:
Displays NO
x
concentration based on the
last attempted background calibration. High alarm shows user
defined limit for acceptable background check offset.
●
NO Span Check Offset:
Displays NO concentration based on the
last attempted span calibration. High alarm shows user defined
limit for acceptable span check offset (compared to the span
concentration value). Span concentration shows span setpoint.
●
NO
x
Span Check Offset:
Displays NO
x
concentration based on the
last attempted span calibration. High alarm shows user defined
limit for acceptable span check offset (compared to the span
concentration value). Span concentration shows span setpoint.
Note
If both the low alarm and high alarms are set to zero, then no alarm
will show.
▲
Summary of Contents for 42iQTL
Page 1: ...42iQTL Instruction Manual Trace Level NO NO2 NOx Analyzer 119019 00 1April2020 ...
Page 2: ......
Page 26: ......
Page 34: ...Operation Instrument Display 3 8 42iQTL Instruction Manual Thermo Scientific ...
Page 162: ......
Page 190: ......
Page 258: ......
Page 292: ......
Page 304: ......
Page 308: ......
Page 309: ......