Servicing
DMC Ozonator
Thermo Scientific
42iQTL Instruction Manual
7-47
Figure 7–47.
HPVS Remove Bd
Use the following procedure to replace the flow switch.
Equipment Required:
Phillips drive, #2
1.
Remove Ozonator as described in “
DMC Ozonator
” in this chapter.
2.
Unplug flow switch (J5) from board.
3.
Pull flow switch off clip after unplugging from board.
Flow Switch
Replacement