Operation
Calibration
Thermo Scientific
42iQTL Instruction Manual
3-25
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Backgrounds>Adjust Prereactor Background
The Adjust NO, NO
x
, and Prereactor Background screens contains the
following information:
●
Adjust NO Background:
User manually adjusts zero NO background.
●
Adjusted NO Concentration:
Read only. Displays what the NO
concentration value will become when the calibrate button is pressed.
●
Current NO Concentration:
Read only. Displays the current NO
concentration.
●
Adjust NO
x
Background:
User manually adjusts zero NO
x
background.
●
Adjusted NO
x
Concentration:
Read only. Displays what the NO
x
concentration value will become when the calibrate button is pressed.
●
Current NO
x
Concentration:
Read only. Displays the current NO
x
concentration.
●
Adjust Prereactor Background:
User manually adjusts zero Prereactor
background.
●
Adjusted Prereactor Concentration:
Read only. Displays what the
Prereactor concentration value will become when the calibrate button is
pressed.
●
Current Prereactor Concentration:
Read only. Displays the current
Prereactor concentration.
●
Calibrate:
When pressed, updates the background.
Summary of Contents for 42iQTL
Page 1: ...42iQTL Instruction Manual Trace Level NO NO2 NOx Analyzer 119019 00 1April2020 ...
Page 2: ......
Page 26: ......
Page 34: ...Operation Instrument Display 3 8 42iQTL Instruction Manual Thermo Scientific ...
Page 162: ......
Page 190: ......
Page 258: ......
Page 292: ......
Page 304: ......
Page 308: ......
Page 309: ......