Calibration
Calibration in Dual Range and Auto Range Mode
Thermo Scientific
42iQTL Instruction Manual
4-19
3.
When the responses stabilize, from the Home screen, choose
Calibration>Calibrate Span Coefficients>Calibrate Low Range NO
Span Coefficient
.
The user sets the span concentration by pressing the Edit Span
Concentration button. The Calculated Low Range NO Span
Coefficient button will show what the span coefficient will be set to if
the Calibrate button is pressed.
4.
Enter the NO calibration gas concentration using the pushbuttons, and
then press
Calibrate
to calculate and save the new low range NO span
coefficient based on the entered span coefficient.
Use the following procedure to calibrate the NO
x
channel to the NO
x
calibration gas.
1.
Press the Back button to return to the Calibration menu and choose
Calibrate Low Range NO
x
Span Coefficient
.
The user sets the span concentration by pressing the Edit Span
Concentration button. The Calculated Low Range NO
x
Span
Coefficient button will show what the span coefficient will be set to if
the Calibrate button is pressed.
2.
Verify that the low NO
x
calibration gas concentration is the same as the
low NO calibration gas concentration plus any known NO
2
impurity.
3.
Enter the NO
x
calibration gas concentration using the pushbuttons,
and then press
Calibrate
to calculate and save the new low range NO
x
span coefficient based on the entered span coefficient.
Use the following procedure to calibrate the NO
2
channel to the NO
2
calibration gas.
1.
Adjust the O
3
generator in the GPT system to generate sufficient O
3
to
produce a decrease in the low NO concentration equivalent to about
80% of the URL of the low NO
2
range. The decrease must not exceed
90% of the low NO concentration determined in the “
Calibrate NOx
Low
” procedure.
2.
Press the Back button to return to the Calibration menu and choose
Calibrate Low Range NO
2
Span Coefficient
.
Calibrate NO
x
Low
Calibrate NO
2
Low
Summary of Contents for 42iQTL
Page 1: ...42iQTL Instruction Manual Trace Level NO NO2 NOx Analyzer 119019 00 1April2020 ...
Page 2: ......
Page 26: ......
Page 34: ...Operation Instrument Display 3 8 42iQTL Instruction Manual Thermo Scientific ...
Page 162: ......
Page 190: ......
Page 258: ......
Page 292: ......
Page 304: ......
Page 308: ......
Page 309: ......