Servicing
DMC PMT Cooler and Reaction Chamber Replacement
Thermo Scientific
42iQTL Instruction Manual
7-29
Use the following procedure to replace the PMT cooler and clean or
replace the reaction chamber assembly.
●
Remove the PMT Cooler
●
Replacing the PMT board
●
Removing and Cleaning the Reaction chamber
●
Replacing the PMT Cooler and Reaction chamber
●
PMT Replacement
Use the following procedure to remove the PMT cooler from the
instrument.
Equipment Required:
Phillips drive, #2
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover (
Figure 2–1
).
Make sure the PMT has cooled down before touching.
▲
2.
Unhook plumbing by pulling loop downwards and unhook tubing.
3.
Using a #2 Phillips drive, unfasten two 6-32 flat head screws.
4.
Pull shroud outwards.
DMC PMT Cooler
and Reaction
Chamber
Replacement
PMT Cooler Shroud
Removal
Icon Here