Operation
Calibration
Thermo Scientific
42iQTL Instruction Manual
3-11
The Calibrate NO, NO
x
, and Prereactor Background screens are used to
calibrate the instrument zero background. Before making an adjustment,
be sure the analyzer samples zero air for at least 5 minutes.
It is important to note the averaging time when calibrating. The longer the
averaging time the more precise the calibration results. To achieve
maximum precision, allow the instrument to stabilize each time input gas is
changed and set the averaging time to 300-second averaging.
Home Screen>Calibration>Calibrate NO Background
Home Screen>Calibration>Calibrate NO
x
Background
Calibrate NO, NO
x
, and
Prereactor Background