Servicing
DMC Ozonator
7-44
42iQTL Instruction Manual
Thermo Scientific
4.
Using a #2 Phillips drive, unfasten captive hardware 4X.
Figure 7–44.
HVPS Remove
Use the following procedure to replace the ozonator.
1.
Unplug ozonator (J1).
2.
Disconnect plumbing 2X.
3.
Ozonator—using a #2 Phillips drive, unfasten two #6-32 pan head
screws.
Ozonator
Replacement