Servicing
DMC Pressure and Flow Board
Thermo Scientific
42iQTL Instruction Manual
7-17
Use the following to replace the DMC pressure and flow board.
Equipment required:
Hex drive, 7/16
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug cables from the pressure and flow board.
Figure 7–20.
Flow Pressure Board, Disconnect DMC Cable
3.
Disconnect plumbing.
4.
Using 7/16 hex drive, unfasten four #6-32 socket cap head screws.
DMC Pressure
and Flow Board