Operation
Calibration
3-26
42iQLS Instruction Manual
Thermo Scientific
The Adjust Span Coefficients screen is used to manually adjust the span
coefficients.
The following screens show the manual calibration screens in single range
mode and dual or auto range mode. In dual or auto range modes, “High
Range” or “Low Range” buttons are displayed to indicate the calibration of
the high or low coefficient. The Adjust High Span Coefficient and Adjust
Low Span Coefficient screens function the same way as the (single range)
Adjust Span Coefficient screen.
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Span Coefficients (single range mode)
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Span Coefficients (dual or auto range mode)
The Adjust Span Coefficients screen contains the following information:
●
Calibrate NO Coefficient:
Adjusts the NO coefficient when in single
range mode.
●
Calibrate NO High Range Coefficient:
Adjusts the NO high range
coefficient when in dual or auto range mode.
Adjust Span Coefficients
Summary of Contents for 42iQLS
Page 1: ...42iQLS Instruction Manual Low Source NO NO2 NOx Analyzer 117436 00 1Sep2019 ...
Page 2: ......
Page 18: ......
Page 34: ...Operation Instrument Display 3 8 42iQLS Instruction Manual Thermo Scientific ...
Page 192: ......
Page 296: ......
Page 304: ......
Page 316: ......
Page 317: ......