Operation
Settings
3-94
42iQLS Instruction Manual
Thermo Scientific
●
Down:
●
Manual Entry:
Shows current value. If the input board cal cell is
pressed, user can manually change the value.
●
Auto Cal:
When pressed, the auto-calibration process is initiated.
Please allow up to 5 minutes for calibration to complete.
●
End Cal:
When pressed, the auto calibration is interrupted and the
value does not change.
●
PMT Voltage:
Displays PMT voltage.
●
Frequency:
Displays frequency.
●
Zero Offset Freq:
Displays zero offsets frequency.
●
Default Values:
When pressed, the default values are saved.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Reaction Chamber Settings>Continue to PMT Calibration
Screen
The PMT Calibration screen contains the following information:
●
Across:
●
Description:
Defines the actions the user can do.
●
PMT Supply:
User can manually adjust the PMT supply value,
adjust target concentration and read calibration concentration,
start/stop automatic calibration, read voltage, frequency and zero
offset frequency values, and reset default values.
●
Current PMT:
Displays the current PMT voltage value.
●
Units:
Displays units for each item.